Graphical user interface simplifies infusion pump programming and enhances the ability to detect pump-related faults.

Drug administration errors are frequent and are often associated with the misuse of IV infusion pumps. One source of these errors may be the infusion pump's user interface.

We used failure modes-and-effects analyses to identify programming errors and to guide the design of a new syringe pump user interface. We designed the new user interface to clearly show the pump's operating state simultaneously in more than 1 monitoring location. We evaluated anesthesia residents in laboratory and simulated environments on programming accuracy and error detection between the new user interface and the user interface of a commercially available infusion pump. The number of programming errors and workload were reduced partly because it took less time and fewer keystrokes to program the pump when using the new user interface. Despite minimal training, residents quickly identified preexisting infusion pump problems with the new user interface. Intuitive and easy-to-program infusion pump interfaces may reduce drug administration errors and infusion pump-related adverse events.

Additional Info

  • Date: 2012
  • Type: Research
  • Investigators / Instructors / Facilitators:

    Syroid N, Liu D, Albert R, Agutter J, Egan TD, Pace NL, Johnson KB, Dowdle MR, Pulsipher D, Westenskow DR.